Maintenance of High Pressure Steam Sterilizer

2020/12/3 10:32:58

1. The high-pressure steam sterilizer regularly disinfects the water storage tank with medical disinfection water every week. First empty the distilled water in the water storage tank, add medical disinfectant water to the water storage tank, let it stand for one hour and then empty it, and then wash it again with distilled water.
2. Use medical alcohol to disinfect and clean the surface of the inner cavity of the high-pressure steam sterilizer every month.
3. Clean the filter. Open the door and twist out the filter counterclockwise, rinse with clean water and use a brush to remove the impurities adsorbed on the surface. If possible, you can also put it in a washing machine for cleaning.
4. Replace the fuse. When the fuse is broken, please turn the fuse holder counterclockwise by hand to remove the fuse after power off. After replacing the fuse of the same specification, turn the fuse holder clockwise by hand.
5. Regularly clean the sealing surface to prevent impurities left on the surface after long-term use and affect the sealing performance. Dip a lint-free cloth with distilled water, and gently wipe the surface of the seal ring or seal cover. After the above treatment, if there is still air leakage, you need to remove the sealing ring for cleaning, if it is damaged, it must be replaced.
6. Replacement of sealing ring
A. When the seal ring is aging and fails, pinch the lips of the seal ring and pull down the whole.
B. Clean the sealing ring installation groove.

C. Gently press the new sealing ring into the sealing ring installation groove until the entire sealing ring enters the installation groove. Do not pry or press it with a sharp object such as a screwdriver. Just press in lightly.